Infant Incubator

Product Description

Servo Control Mode:

air temperature, skintemperature, display humidity;

Parameter Displaying:

set temperature,air temperature, skin temperature, internal humidity, heat power rate;

Audible & Visual Alarms:

powerfailure, sensor failure, higher & lower temp, air & skin temperature override,fan failure, system failure;

Preset Mode:

power failure reset memory, alarm silence, keypad lock, self-testfunction;
Hood with dual slope; 6 hand-ports;
RS-232 Output Connector (use for networksystem);
IV poles;
Protective guard rail;
Storage cabinets.

Specifications of hood

Mattressto Hood: 450 mm
Observationside of hood: Two incline
Operateport: 6 (2 Iris ports)
Tubinggrommets: 4
Frontaccess panel: 826×310 mm
MattressTray Size: 630×360 mm

Technical Specifications:

Powerrequirement ≤400VA
Temperaturecontrol mode: Air and skin servo control
AirTemp Control Range: 20℃~37℃
Temperatureoverride mode to: 20℃~39℃
SkinTemp Control Range: 30℃~37℃
Temperatureoverride mode to: 30℃~38℃
Skintemp sensor accuracy: ≤0.3℃
Humiditydisplay range: 5%~99%
Temperaturefluctuation: ≤0.5℃
Mattresstemperature uniformity: ≤0.8℃
Temperaturerise time: ≤45 min
Noiselevel within hood environment: ≤55dB

Alarms

Powerfailure alarm
Fanfailure alarm
Sensorfailure alarm
Highair temperature alarm: +3.0℃
Low airtemperature alarm: -3.0℃
Highskin temperature alarm: +1.0℃
Lowskin temperature alarm: -1.0℃
Overair temperature alarm: ≤38.0℃; ≤40.0℃
Overskin temperature alarm: ≤38.0℃; ≤39.0℃
Systemfailure alarm
Alarmsilence function

Operational condition

Environmenttemperature: +20℃~+30℃
Environmentrelative humidity: 5%~99%
Environmentair velocity of flow: <0.3m/s

Others

Infantbed tilt adjusting ±10° step-less
Mattresstray Max. load 10kg
Transfusionpole Max. load 2kg
CabinetMax. load 6kg